Intech Auto TLD Badge Reader
- Improved, smooth & noise-less mechanical transport design for trouble free service.
- Linear actuators for raising card and for PMT shutter
- Interchangeable magazines.
- More efficient Hot gas heater design – fast heating & cooling
- Fast magazine ejection with two speed drive for magazine movement
- Environmentally safer temperature compensated blue LED light source
- Peltier cooling for PMT for better stability of background
- USB 2.0 serial interface
Improved, wider range I-F converter with better linearity from 0.1nA to 100µA to cover the range of 0-10Sv
SPECIFICATIONS
Dosimeter Badge : Three-element BARC CaSO4(Dy) PTFE disc dosimeter badge with machine readable 7 digit ID
Light Measurement System : Photomultiplier tube ( EMI 9125A – bialkali ) Light measuring system (LMS) with integrated HV and I-F Converter
Dark current : Dark current is 1 Sv (CaSO4) equivalent with improved software based sampling & subtraction
Heating Method : Hot gas (N2) heating
Heating Cycle : The temperature is raised to 280C in 8 sec and clamped at 280C
Dose Range Detection Threshold
50 Sv – 10Sv (gamma) < 20 mSv.
100Sv – 10Sv (beta)
Readout time Residual signal
100 secs. per badge < 10% of reading
Facilities Available
- Entry of badge ID Nos, calibration factor, etc.
- Storage of dose and glow curve data of badges.
- Motorised driver assembly for automatic feeding of 50 dosimeter cards loaded in a magazine. Software
- Windows Xp, Windows 7, compatible, menu driven software for transfer of data to a user defined file, storage & display of glow curves, computation of dose & generation of dose reports.
Temperature Monitoring
Chromel Alumel thermocouple in hot gas stream.
Range Selection
Automatic from 9.999mSv to 9.999Sv
Calibration
By adjustment of EHT through a potentiometer in the EHT circuit
Safeguards
Heater / Gas flow Failure: The heater and gas flow are checked for failure in every dosimeter readout cycle. In the event of failure of heater or gas flow the readout is terminated and a message indicating heater / gas flow failure is flashed on the PC monitor.
Mechanical failure: Any mechanical failure during readout cycle is sensed using a time-out watch dog programme and the cycle is terminated with an option for the user to restart the cycle.
EHT/ Input circuit failure: Failure of EHT or the input circuit (I-F converter) is sensed and reading cycle is terminated in case of failure.
Power
Power supply :230V, 50 Hz : power requirements: 500 VA (max including PC)
PC requirements
Intel Core i3 – 2120 or upward compatible, with licensed Windows 7 OS with standard peripheral devices.
Operating System